@Article{JFBI-16-381, author = {Kou , Tian-Yu and Li , Henry Yi}, title = {Factors Influencing the Accuracy and Precision of NFC-Based Temperature Monitoring Device}, journal = {Journal of Fiber Bioengineering and Informatics}, year = {2024}, volume = {16}, number = {4}, pages = {381--396}, abstract = {
Near-field communication (NFC) is a short-distance wireless data transmission technology with potential for wearable sensors. Xu et al. (2020) developed a battery-free smart textile patch with an NFC antenna and a temperature sensor (STP-NFC) [1]. To address potential issues with the testing protocol that could affect the accuracy and precision of temperature measurement from the STP-NFC, a validation experiment was conducted at a fixed room temperature of ${\rm 20}^◦{\rm C}.$ This was done using a commercial IR imager to compare the results with the STP-NFC results. Results showed excellent accuracy with an average temperature of ${\rm 21.50}^◦{\rm C}$ and a difference of only ${\rm 0.36}^◦{\rm C}$ from an IR imager’s reading. The STP-NFC also had excellent precision with a small standard deviation of 0.83. Optimal performance was achieved with a 6 mm distance, a 150-second time interval, and a 4-second scanning duration for each scan. Additionally, the gage repeatability and reproducibility (R&R) study has been conducted to assess the STP-NFC measurement system’s consistency, and the STP-NFC’s reproducibility has been demonstrated. These results have implications for developing reliable wearable medical monitoring devices using NFC technology.
}, issn = {2617-8699}, doi = {https://doi.org/10.3993/jfbim02571}, url = {http://global-sci.org/intro/article_detail/jfbi/23433.html} }